Design of an Electrooculography Equipped with Digital Neck Angle Elevation Gauge

Authors

  • Husein Mahendra Department of Medical Electronics Technology, Poltekkes Kemenkes Surabaya Author
  • Endro Yulianto Department of Medical Electronics Technology, Poltekkes Kemenkes Surabaya, Surabaya, Indonesia Author
  • Muzni Zahar Politeknik Caltex Riau, Riau, Indonesia Author

Keywords:

Incubator Analyzer, Temperature, Humidity, TFT Display

Abstract

Human eye health problems due to uncontrolled use of gadgets can cause eye fatigue. However, medical devices for diagnosing the electrical activity of the eye muscles have received less attention. This study designed a tool that can monitor the electrical eye muscles (EOG) equipped with a neck angle elevation meter, to see if there is a difference in signal when someone is working in front of a computer screen at a certain angle elevation. This study designed an EOG module consisting of an instrument amplifier, a pre amplifier circuit, a high pass filter, a low pass filter, a summing amplifier, and a notch filter. In this study, the lead points are at the vertical point and the horizontal point of the eye using disposable electrodes. The frequency of the signal used is 0.5-30 Hz. The output results were compared with KandH as a standard tool. . In this study, the results obtained the smallest error on the vertical channel of -1.35622% and on the horizontal channel of -4.32745%. For conditioning carried out on respondents, all respondents experienced a decrease in amplitude with an elevation of the neck angle of 60°. Thus, it can be concluded that the elevation of the neck angle affects the electrical power of the eye muscles. For further development, an alarm or eye fatigue indicator can be added.

Downloads

Published

30-05-2025

Issue

Section

Articles